gvtcx

Dilution Refrigerator for Cryogenic Characterisation of RF Hardware

NPL Management Limited Published 20 Apr 2026 Find a Tender

key details

Value£500,000 (£600,000 inc. VAT)
Statusplanning
Category (CPV) 31000000 +2 more
SME suitableYes
OCIDocds-h6vhtk-0685fe

description

NPL is seeking engagement from suppliers who can provide "Dry" pulse tube cooled dilution refrigerator system to be used for characterisation of quantum hardware including RF and quantum circuits with sample volume under vacuum. The system should include thermometry and associated wiring and readout electronics with provision to integrate additional temperature sensors in future for comprehensive diagnostics. The purpose is to complement our existing measurement systems and add measurement capability, for characterisation of quantum and RF hardware with high throughput and faster cooldowns repeated typically every 2 weeks.

The dilution refrigerator needs to be suitable for RF measurements of complex quantum circuits and hardware, meaning it must be equipped with 20+ coaxial lines with provision to further increase the number of coaxial lines in the future. These lines must include high attenuation input lines suitable for use with sensitive quantum devices such as superconducting qubits and parametric amplifiers and low-loss output coaxial lines suitable for use with low-noise amplifiers, isolators and parametric amplifiers. Some of these coaxial lines must operate up to 40 GHz with provision to integrate even higher frequency RF cabling in the future. There should be sufficient number of DC lines with sufficient reconfigurability to enable biasing of at least tens of RF multi-way switches, at least four low noise amplifiers and few dedicated high resistance and low resistance DC looms for devices under test with provision to integrate more DC looms in the future.

The asset must enable the characterisation of quantum and cryogenic RF hardware based on the following:

1)Base temperature of less than or equal to 10 mK with access to PT2 and other intermediate stages for testing components at intermediate temperatures.
2)Mixing chamber plate diameter of at least 500 mm and vertical experimental sample space below the mixing chamber with a height of at least 400 mm accessible for custom experimental apparatus
3)We do NOT require a magnet.

This is not an evaluation exercise and nothing in this activity will limit the opportunity to engage in any follow on ITT.

notice history

1 notice published against this procurement.

PublishedTypeRegimeNotice
20 Apr 2026 Preliminary market engagement (UK2) Procurement Act 2023 035497-2026

more from NPL Management Limited

all contracts from this buyer →

similar contracts awarded

Other awarded contracts in the same category. Useful for seeing who normally wins this kind of work, and at what value.

source

Published on Find a Tender. Contact details for named individuals are not reproduced on this site.