CEFAS25-134 Request for Quotation for Acquisition of an ICP-OES

CEFAS Published 10 Feb 2026 Contracts Finder
The value below is a framework or dynamic market ceiling: the maximum that could be spent across all call-offs, not the value of a single contract. Aggregate figures on this site exclude these to avoid double counting.

key details

Value£51,453
Statuscomplete
Category (CPV) 38433000
Deadline8 Jan 2026
Contract start10 Feb 2026
Contract end10 Feb 2027
Procedureselective
SME suitableYes
OCIDocds-b5fd17-75ddf1e4-49e1-4a3d-8e79-7510420f167d

Award

SupplierValueDateStatus
Thermo Fisher Scientific £51,453 9 Feb 2026 active

description

The Chemistry Function at the Cefas Lowestoft Laboratory has a requirement for an inductively coupled plasma optical emission spectrometer (ICP-OES) for the analysis of environmental samples. The system will be capable of analysing trace elements in diluted acid matrix solutions (HNO3, HCl and HF) formed from microwave acid digestion of marine sediment and biota samples. Ex-demonstration instruments under 3 years old may be proposed at a reduced price and provided 12 months extended warranty is offered. Where an ex-demonstration instrument is proposed, this must be clearly declared in the supplier's quotation.

This purchase is being conducted under the LUPC Framework Agreement: High Value Laboratory Equipment (HVLE) - LAB3162 NW, Lot 3 - Spectroscopy Equipment.

Scope of Requirements:

  1. Technical Requirement

The requirement is for a fully automated, bench-top ICP-OES system capable of high-throughput, multi-elemental analysis with robust performance for a wide range of sample matrices. The system must deliver high sensitivity, low detection limits, and excellent long-term stability, supporting both routine and research applications. The objective is to enhance analytical capacity, improve data quality, and ensure compliance with relevant standards (e.g., ISO 17025).

The OES system must:

  • Have an argon gas source which it uses efficiently.
  • Be axial mode (or duo mode), and not purely radial mode.
  • Have sub-1ppb detection limits for most required elements (at optimal wavelength).
  • Operate under a wavelength range of approximately 180 to 850 nm or better.
  • Be robust and demonstrate minimal downtime due to maintenance.
  • Include the chiller unit.
  • Include an autosampler and associated paraphernalia.
  • Include a sample tray containing at least 150 positions (with options of rack configuration for standard and sample solutions).
  • 2.Environmental Considerations
  • Unit must have an ultra-low power consumption.
  • Unit must be designed for reuse/remanufacture and of regenerative nature to eliminate waste where possible.
  • Consideration of how packaging will be managed. Ideally at the time of installation the supplier will remove packaging and dispose of by ethical and environmentally friendly methods. Where possible packaging is to be re-used or recycled.
  • Consideration of life cycle assessment for the equipment.
  • Trade-in or removal of the old system will be of benefit. If a trade-in option can be provided, please include the relevant details in the Pricing Schedule 2 table of this RFQ.

notice history

1 notice published against this procurement.

PublishedTypeRegimeNotice
10 Feb 2026 Award (award) · a474130d-ed1e-4358-bc01-f2ec810d76cf-882875