gvtcx

Benchtop Nuclear Magnetic Resonance System

University of Huddersfield e-Tendering Published 21 Sept 2017 Contracts Finder

key details

Statusactive
Category (CPV) 33113110 +3 more
RegionYorkshire and the Humber
Deadline6 Oct 2017
Contract start13 Oct 2017
Contract end31 Jan 2018
Procedureopen
SME suitableNo
OCIDocds-b5fd17-6ca204b1-16ab-437f-9c14-3b611456c3ad

description

The University of Huddersfield seeks to purchase a Benchtop NMR system for optically generated hyperpolarization studies.

A benchtop nuclear magnetic resonance (NMR) system is required for research into a new method of signal enhancement in NMR spectroscopy, which is dependent upon optical illumination of the liquid sample prior to measurement. The research project will include an investigation of the effects of illumination at different magnetic field strengths prior to transfer to the detection field.

The key requirement of the system is therefore that optical access must be possible to illuminate the sample at a variable height within the magnet bore where the static magnetic field is ≥ 350 mT, and that rapid flow from the illumination region to the NMR detection coil is possible (transfer time to be short compared to T1, maximum transfer time 0.5 s). Minimum permissible clearance around sample tubing in magnet bore for insertion of fibre optics is 6 mm radially.

NMR spectrometer specifications:

1H and 13C detection at a magnetic field of around 1 T (approx. 43 MHz and 11 MHz NMR frequencies)

High 1H sensitivity (at least 50:1 in single scan on 1% ethyl benzene or similar)

Linewidth < 0.5 Hz at 50% and < 20Hz at 0.55%

Field stability better than 0.01 ppm

Deuterium free lock for experiments on non-deuterated solvents

Ability to shim on user sample in addition to any manufacturer specified standard sample

Demonstration that shimming is stable during laser illumination (min. 1 W at 532 nm) for up to 10 s within magnet bore

Desirable: Magnet temperature readout to indicate if laser illumination has caused magnet heating

Ability to write user-defined pulse programmes or scripts triggering a minimum of two separate TTL outputs for synchronisation with laser equipment at a defined time relative to NMR acquisition. Low jitter level for TTL outputs required, 0.1 s max.

Available pulse sequences to include standard 1D and 2D experiments for each nucleus, reaction monitoring, relaxation measurements (T1 and T2), and 13C detection with 1H decoupling.

Flow system:

Pump and all connecting tubing supplied for continuous or intermittent flow of sample solutions through spectrometer

Spectrometer stand to be supplied to allow easy access for insertion of optical components into a vertical magnet bore from below

Flow system made from optically transparent (T > 90% 400-700 nm) material (e.g. glass or quartz) in the accessible illumination region (≥ 350 mT) within the magnet bore

Computer control of sample flow for synchronisation with user-defined NMR pulse programmes or reaction monitoring scripts

System requirements:

Cryogen-free system

No external cooling water needed

Computing:

Spectrometer to be supplied with control computer and all standard peripherals (monitor, keyboard, mouse); minimum hard ...

Additional information:

Please apply for this tender through our In Tend platform which is available through the below link

https://in-tendhost.co.uk/universityofhuddersfield/aspx/Home

Is a Recurrent Procurement Type? : No

documents

Documents are linked, not mirrored. They are served by the publishing authority and may require registration.

notice history

1 notice published against this procurement.

PublishedTypeRegimeNotice
21 Sept 2017 Tender (tender) · e05ff267-9fc8-4d44-a051-fd3ebef32060-171950

more from University of Huddersfield e-Tendering

all contracts from this buyer →

similar contracts awarded

Other awarded contracts in the same category. Useful for seeing who normally wins this kind of work, and at what value.

source